Quantitative scale of radiodensity From Wikipedia, the free encyclopedia
Remove ads
The Hounsfield scale (/ˈhaʊnzfiːld/ HOWNZ-feeld), named after Sir Godfrey Hounsfield, is a quantitative scale for describing radiodensity. It is frequently used in CT scans, where its value is also termed CT number.
Definition
Summarize
Perspective
The Hounsfield unit (HU) scale is a linear transformation of the original linear attenuation coefficient measurement into one in which the radiodensity of distilled water at standard pressure and temperature (STP) is defined as 0 Hounsfield units (HU), while the radiodensity of air at STP is defined as −1000 HU. In a voxel with average linear attenuation coefficient , the corresponding HU value is therefore given by:[1][2]
where and are respectively the linear attenuation coefficients of water and air.
Thus, a change of one Hounsfield unit (HU) represents a change of 0.1% of the attenuation coefficient of water since the attenuation coefficient of air is nearly zero.[3]: 259
Calibration tests of HU with reference to water and other materials may be done to ensure standardised response. This is particularly important for CT scans used in radiotherapy treatment planning, where HU is converted to electron density.[4] Variation in the measured values of reference materials with known composition, and variation between and within slices may be used as part of test procedures.[3]: 283 [5]
Rationale
The above standards were chosen originally to encode the radiodensity of organic tissues relative to water for 12-bit processing on clinical CT scanners.[6] A 12-bit encoding corresponds to 4096 () values, where the range (–1024 to 3071) encompasses HU values for air, soft tissue and bone.
Remove ads
Values for different body tissues and material
Summarize
Perspective




HU-based differentiation of material applies to medical-grade dual-energy CT scans but not to cone beam computed tomography (CBCT) scans, as CBCT scans provide unreliable HU readings.[7]
Values reported here are approximations. Different dynamics are reported from one study to another.[citation needed]
Exact HU dynamics can vary from one CT acquisition to another due to CT acquisition and reconstruction parameters (kV, filters, reconstruction algorithms, etc.). The use of contrast agents modifies HU as well in some body parts (mainly blood).
A practical application of this is in evaluation of tumors, where, for example, an adrenal tumor with a radiodensity of less than 10 HU is rather fatty in composition and almost certainly a benign adrenal adenoma.[31]